CerebroGate: Redesigning the Ventricular Catheter
Program:
Biomedical Engineering
Project Description:
Hydrocephalus is a condition characterized by accumulation of excess cerebrospinal fluid (CSF) in the brain’s ventricles. The most common treatment method for hydrocephalus is a shunt device with three main components: a ventricular catheter, a shunt valve, and a distal catheter. The issue with shunts is that 98% fail within 10 years, with the largest source of malfunction being ventricular catheter obstruction. Ventricular catheter obstruction can be caused by many different modes, including choroid plexus entanglement, astrocyte growth, blood clots, and protein clots. However, there currently exists no effective, validated model of ventricular catheter obstruction that can incorporate these different modes. Our poster details the development of a ventricular obstruction model for improved catheter testing. Our model accurately replicates physiological fluid flow conditions with inflow and outflow of 0.6 mL/min, incorporating protein and blood obstructions to mimic real-world scenarios.
Team Members
-
[foreach 357]
[if 397 not_equal=””][/if 397][395]
[/foreach 357]
Project Mentors, Sponsors, and Partners
Kurt Lehner, MD
Course Faculty
-
[foreach 429]
[if 433 not_equal=””][/if 433][431]
[/foreach 429]
Project Links
Additional Project Information
Project Photo
Project Photo Caption:
The image shows a top view of our ventricular obstruction model with its three main components: a peristaltic pump, a ventricular box, and a control ventricular catheter.